The Automated Nucleic Acid Extraction Workstation SAW-96 (referred to as "SAW-96" below) produced by Jiangsu Bioperfectus Technologies Co., Ltd. is an in-vitro diagnostic medical device that integrates cutting-edge mechanical, electronic, and software technologies. SAW-96 integrates sample loading, nucleic acid purification, and PCR setup functions in one instrument. The typical run duration is less than 70 minutes for 1-96 samples. Based on experienced magnetic separation technology, SAW-96 provides high throughput automation combined with high performance. It will promote laboratory productivity to meet changing needs. With non-contact dispensing of samples, reagents, and PCR reagents, it will be safe and simple for the operators and save time and money for the laboratory.
Scope of application: The equipment is intended for automatic extraction workflow. It can realize automated operations from primary sample tubes to nucleic acids extraction and PCR setup.
Size (L x D x H) | 1,130 x 680 x 690 (mm) |
Weight | 110.8 kg |
Sample throughput | 1-96 |
Cooler control | Keep 4 - 15 °C for the reagent station and 4 - 8 °C for the PCR setup station. |
Contamination control |
The nucleic acid extraction and solution dispensing operations are performed strictly in separate areas. A built-in medical purifier consists of two UV disinfection modules with a timer, two HEPA filters, and one high efficiency & high voltage electrostatic sorption particulate air filter with four fans. |
Voltage | 100-240 V AC |
Frequency | 50-60 Hz |
Rated power | 1,210 VA |
Sampling Function | |
Compatible sample tube | Sample tubes and centrifuge tubes, such as 1.5 ml tubes, 5 ml tubes, and tubes with 16mm diameter. |
Sample Types | Plasma, serum, whole blood, swab, urine, etc. |
Extraction Function | |
Magnetic bead recovery efficiency | ≥ 98% |
Processing volume | 20-1,000 μL |
Mixing mode | Oscillating mixing with multiple modes and speeds (> 20 gears) |
Compatible kit | Bioperfectus magnetic bead-based nucleic acid extraction kits |
Heating temperature |
Heating temperature for lysis: room temperature to 120°C |
Heating temperature for elution: room temperature to 120°C |
|
Pipetting Function |
|
Sample Loading Channels |
1-2 |
Reagent tube |
1.5 ml, 2 ml, 5 ml EP Tubes 0.2 ml PCR tube |
Dispensing volume of plunger pump |
1-250 μL |
Positional accuracy |
±0.1 mm on the X-Y-Z axis |
NMPA, CE, FDA Listed
I am interested in more information on the following Bioperfectus product